Photonic Instruments
Search

MicroPoint Manual

Hardware Specifications

Model

The MicroPoint system is available optimized for a factory set range of focus.  The angular and spatial alignment of the pulsed dye laser illumination is controlled via 2-axis joy stick.  The focus in the z-direction is controlled with a knurled focus ring.

Model 2200 2203 2204 2205
Range of focus (mm) -300 to -150 -38 to +500 +38 to +∞ -30 to +400

 

Tunable Fiber Optic Dye Pumped Source

Average power 750microWatts @ 15Hz
Peak power 12kW
Pulse energy 50microJoules
Stability +/- 3%
Pulsewidth 3 to 5nsec
Pulse repetition rate 0 to 15Hz
CDRH IIIb

 

Optical

Transmission 365nm to 700nm
Spectral bandwidth 4nm FWHM (typical)
Attenuation options Manual, 28 steps, 0.1%-100% transmission
Motorized, 90 steps, 0.1%-100% transmission
Resolvable spot size Near diffraction limited

 

Mechanical / Electrical

Illumination port clear aperture Ø22mm or Ø34mm
Illumination port filter size Ø25mm or Ø38mm
Dimensions 8.5" (l) x 6.0" (w) x 2.75" (h)
10.0" (l) x 6.0" (w) x 3.75" (h)
Weight (laser head) 3 pounds
Lifetime 20,000,000 laser pulses
30,000 laser pulses per refillable dye cell

 

Illumination Port Options

Dichroic beamspliter Single pass, specify wavelength
Multi pass, specify wavelength
Beamsplitter 450nm to 750nm, R = 100, 70, 50 or 30%
Excitation filter 360nm / 40nm (DAPPI)
480nm / 20nm (GFP)
470nm / 40nm (FITC)
535nm / 40nm (Rhodamine)

Additional illumination port options on request

 

Adapters

Flange adapters are available to retrofit to the epi-port or video port of commercially available microscopes.

Flange Part No.
Leica 2237
Nikon 2235
Olympus 2236
Zeiss 2238
C-mount video port 2230C

MODULAR PLATFORM

  • Compatible with commercial microscopes
    • Point & slit scan confocal
    • Spinning disk confocal
    • Wide field
  • Lowest cost to add wavelengths in the field for molecularly selective illumination and ablation - change a dye cell and tune system within the visible spectrum
  • Field installable upgrade option for computer control
  • Custom systems available for OEMs of fluorescence based imaging and laser ablation equipment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Laser Safety Label 
Laser Safety Label